Description
Liya Test Magnesium sulphate and sodium sulfate tests are carried out to determine the strength and durability of aggregates to be used in concrete and to be subjected to repeated freeze-thawing. The left column contains only the specific products used in the experiment. In addition to these products, oven, sieve, scales, etc. Hardware are also required for the experiment.
According to ASTM standards, sodium sulfate may also be used instead of magnesium sulfate. When sodium sulfate is used in the assay, a Hydrometer (1100-1200 g / ml) is required.
The temperature of the test solution is automatically adjusted by the user to the desired temperature by a microprocessor-type sensor, which is placed in sealed containers or stainless steel immersion containers. The temperature of the test solution is displayed in real time on the digital display with a resolution of 0,1 ° C.
Magnesium Sulphate / Sodium Sulphate Test Water Bath has an accuracy of (20 ± 2) 0C. When the immersion cabinet is removed, the water bath can be used as a general purpose water bath at temperatures between 15ºC and 60 ºC.
Supplied with immersion container and lid.